Chapter 4: Running Samples 129
Figure 4-10 Draining liquid from the trap
3 Empty the waste container according to your standard laboratory
procedures for biohazardous waste.
4 Add approximately 1 L of bleach to the waste container (10-L container).
Add a sufficient amount so that 10% of the total waste volume is bleach.
5 Replace the waste trap and attached filter cap. Hand-tighten the trap and
cap until they are fully closed.
6 If one month has passed since you last changed the cap, replace the filter
cap with a new one. Write the date on the new cap as a reminder.
7 Reconnect the sensor and fluid line connectors to their respective ports.
To p revent over-pressu riza tion during fluid ics start up, do not ov erti ghten
the trap or attached filter cap. Tighten each component only until it is
hand-tight. Do not use sealants such as Teflon® tape or other adhesives.
drain plug
trap
338677 Rev A
Waste (A)
space for date