96 BD FACSAria II User’s Guide
The new sheath pressure is automatically saved with the current Sort Setup
mode.
; Tip The name of the configuration is displayed at the top of the
BD FACSDiva workspace. If you use a descriptive name for each
configuration (such as 100-15-Custom, for nozzle size-sheath pressure-Sort
Setup mode), it will be easier to verify that the appropriate settings match
without going back to view the configuration.
Configuration Mismatch Dialog
If there is a mismatch between the CS&T settings and those in BD FACSDiva
software, a dialog opens (Figure 3-7 on page 97) to indicate what the differences
are, and gives these options:
•Use CS&T Settings—Select this option if you have run a performance check
or created a new configuration, and want to use those values.
• Keep Current Settings—Select this option if you were viewing a
configuration in CS&T, and don’t want to change the values. For example,
settings for area scaling for the particle of interest.
Enter the same sheath pressure that was
entered in the configuration.